Location: GoldTech Components Co.,Ltd. » News » MRI scans are getting sharper

News

    Contact Us

    MRI scans are getting sharper

    Guest columnist Claude Gudel, a senior engineer at LEM, believes that the development of fluxgate current transducers can lead to sharper MRI images

    MRI is a powerful medical technology that has revolutionised diagnosis of a very wide range of illnesses and injuries, greatly reducing or in many cases eliminating the need for exploratory surgery. It provides medical practitioners with two- and three-dimensional images, as well as high-accuracy cross-section, of internal structures and organs within a patient’s body.  Underpinning the results achieved by MRI scanning is a wide range of advanced technologies, including precision measurement techniques: the almost unbelievable sharpness of the pictures that MRI produces depends directly on measurements of basic electrical parameters revealing details of soft-tissue structures.

    The working principle is based on nuclear magnetic resonance detecting the magnetic resonance of the protons of hydrogen atoms contained in the water within the human body: water represents up to 70% of body weight. MRI observes the response of the hydrogen nuclei exposed to excitation by both magnetic and electromagnetic fields.

    The collected energy per volume element (voxel) depends on the water distribution in the place under analysis. So MRI can provide a three-dimensional image of the water distribution inside the human body. As each type of body tissue has a characteristic proportion of water within it, it becomes possible to image those tissues, and any deterioration, by looking at changes in water distribution.

    Applying the magnetic fields

    The Static magnetic field Ho, must be very intense, with very high stability and homogeneity within the volume inside the aperture of the MRI scanner, where the patient lies.

    Most of today’s MRIs generate the static field by means of a superconducting magnet located around the cylinder of the scanner. The coils of the magnet are made up of niobium-titanium (NbTi) wires immersed in liquid Helium at a temperature of 4K.
    The Gradient coils superimpose a magnetic gradient to Ho in order to provide a spatial coding of the image. Imaging takes place only in just one plane or slice at a time, and to ensure that signals are received only from nuclei in that plane, only those nuclei have to be pushed to resonance.

    The appearance of the resonance is strongly dependent on the value of the magnetic field Ho: the gradient coils superimpose a magnetic field to ensure that the final magnetic field is exactly equal to Ho only in the plane of interest.

    How the gradient coils work

    To create a gradient along an axis, a pair of coils is needed. In each pair, currents flow in opposite directions (the principle is shown in Figure 1).

     

    Figure 1: Gradient coils add to the static field at one end and diminish it at the other, controlling the plane in which the total field has exactly the correct value.

    In fact, 3 pairs of gradient coils are located around the cylinder of the MRI apparatus to create 3 orthogonal magnetic fields. So, it is possible to adjust the magnetic field at any point in the volume of the cylinder. Gradient amplifiers operating in a closed servo-type loop drive the currents in the gradient coils (Figure 2). Each MRI therefore needs three such current control loops.

     

    Figure 2: Feedback from output current transducers is fundamental to obtaining the required degree of precision from the gradient current amplifier

    The quality, the clarity and resolution of the images are directly linked to those of the magnetic field applied, and therefore to those of the current injected into the gradient coils. One of the key elements in the current control loop is the global accuracy of the current transducer.
    In particular, the following parameters of the current transducer are critical :

    • Extremely low non linearity error (< 3 ppm of measuring range)
    • Very low random noise (low frequency noise from 0.1Hz to 1kHz)
    • Very low offset and sensitivity drifts over temperature range (<0.3 ppm/K)
    • Very high stability of offset versus time (one reason for this is the duration of MRI scans, some of which may last several tens of minutes
    • Measuring range (around 1000 A peak)
    • Bandwidth (–3dB point of 200 kHz)

    To reach these performance levels, Hall Effect current transducers – which were used in previous generations of MRI scanners – are no longer adequate. The solution developed by LEM, primarily for this application area, has similarities to the Hall Effect technique but offers significant advantages. It is described as a double fluxgate, closed loop transducer and identified as type ITL 900. Although fluxgate technology has been available for some time, LEM was able to take this technology and adapt and improve it.

    As well as precise current control in gradient amplifiers for medical imaging, the ITL 900 is equally applicable to measuring feedback in precision current regulated power supplies, current measurement for power analysis, calibration equipment for test benches, and laboratory and metrology equipment which also require high accuracy.

    In its present form, the technology is limited to a relatively narrow operating temperature (typically +10 deg C to +50 deg C).  LEM is confident that the technology can be developed further and the ITL 900 transducer could prove to be as significant for the future of MRI scanning as the Hall Effect transducers were for its introduction. As with Hall Effect itself, with its leading edge performance ITL900 could possibly enable any number of future applications.